Zero point calibration PAN 600AD+
As a rule, the device displays “- – – -” after switching on. However, in some cases a numerical value is displayed even without a measurement.
This value can distort the measured value. In this case, you can correct the measured value as follows:
You can Display value after switching on the measurement result deduct or
You are conducting a Zero point calibration as follows:
Hold down the “HOLD” button (do not release the button), turn on the meter and press the “NCV” button —-
the meter switches to calibration mode. Now release the "HOLD" button, press "NCV" again, the meter beeps, close
the test leads briefly, press the “HOLD” button to start the calibration.
The meter will beep again when calibration is complete.